Logo

GENESYS 30 VISSIBLE SPECTROPHOTOMETER.





GENESYS 30 VISSIBLE SPECTROPHOTOMETER ( PENGGANTI GENESYS 20 )

Product Features:

Entirely new industrial design for better usability
High definition 5-inch color screen with rubberized push-button keypad for easy navigation
Single cell measurement with removable, washable sample compartment liner
Cell holders for standard cell types including test-tubes
New onboard software UI; Thermo Scientific™ VISIONlite PC software is optional
Data any way you want it
USB data export in a text format that’s easy to read and process
Excellent snap-on thermal printer


Specifications :
Baseline Flatness <0.003A 325nm – 1000nm smoothed
Compartments Accessible from top, front or side
Able to accommodate cells up to 100mm pathlength
Available accessory for test-tubes up to 25mm diameter
and 150mm height
Removable, washable sample compartment liner with
magnetic placement and hold-down
Connections USB-A on front face supports flash memory devices for
method and data storage
USB-B on side supports connection to a Windows™
computer running optional remote control software.
Detector Type Silicon Photodiode
Dimensions (L x W x H)35.5cm x 38.5cm x 19.5cm
Display 5" diagonal, 32 bit color display, 800 x 480 pixels
Drift <0.002A/Hr At 500nm after 2hr warmup.
Typical <0.004A/hr after 1hr warmup
Electrical External AC to DC converter.
Voltage and Hz selected automatically, 100-250 volts, 50-60Hz
Keypad Tactile rubber 23 keys with numeric keypad
Lamp 1000 Hrs typical service life
Languages English, Spanish, German, French, Italian,
Portuguese, Russian,Chinese, Japanese, Korean, Thai, Arabic
Noise ≤0.001A at 0A
≤0.001A at 1A
≤0.002A at 2A
RMS at 500nm. 60 consecutive measurements
Optical Design Single beam
Photometric Accuracy ±0.002A (0 - 0.3A)
Instrument 0.5% of ABS reading (0.301A - 2.5A)
Photometric Display -3A to 3A, 0 to 200 000 %T, 0 to 9 999 999 C
Photometric Range -3A to 3A
Photometric ± 0.002A
Repeatability Measured at 1.0 A @546nm
Printer Thermal printer accessory with 80mm width paper
Spectral Bandwidth 5 nm
Stray Light <0.1%T at 340nm and 400nm
Measured using glass cutoff filters or NaNO2 at 340nm
Wavelength Accuracy ± 2nm
Wavelength Data
Interval 1nm
Wavelength Range 325nm – 1100nm
Wavelength
Repeatability < ±1nm
Wavelength Scan
Speed Automatic – up to 1200nm/min
Weight (Metric) 7.5kg
ADDRESS
Rukan Mitra Matraman Blok B - 12
Jl. Matraman Raya No. 148 Jakarta 13150
TELEPHONE
(021) 859 181 54, (021) 859 181 55, (021) 859 023 91, (021) 851 259 5
HANDPHONE
0813 988 55242, 0877 811 22777, 0812 864 04835, 0812 863 8194
FAX
(021) 859 181 78
EMAIL
1. sales@octaprimalestari.com 2. e_anasthasia@octaprimalestari.com, 3. meliati_layarda@octaprimalestari.com 4. octaprima_lestari@yahoo.com
Copyright © 2016 - PT Octa Prima Lestari,
Jasa Pembuatan Website by IKT